Generator of mercury calibration gas MC 3000
The MC-3000 generates a continuous stream of a mercury vapor loaded gas in order to check or calibrate mercury analyzers
An exactly adjustable carrier gas stream (normally air or nitrogen) is loaded with mercury in a first step.
In a second step the gas is supersaturated by cooling. Subsequently the excess mercury is condensed in a special cell to get exact saturation. The gas is then diluted, the dilution ratio is continuously adjustable in order to obtain concentrations within a specific range.
FEATURES & BENEFITS
- Generates elemental mercury vapor
- Continuous operation
- Approx. 10 to 500 µg/m³ Hg°
- Compact sized and portable
- No permeation principle – no need to calibrate the calibrator
- Microprocessor controlled
MAIN APPLICATIONS
Suitable for all applications where a gas stream with a preset and constant mercury concentration is needed
TECHNICAL SPECIFICATIONS |
Range of concentrations |
approx. 10 to 500 µg / m³ Hg° |
Generated calibration gas flow |
1 to 9 l/min, depending on conc. |
Carrier gas |
Air, Nitrogen |
Carrier gas flow |
max. 12.5 l(min) |
Carrier gas pressure |
1 to 3 bar |
Cleansing of carrier gas |
builtin particular filter (0.2 µm ) |
Mercury absorption filter |
sulphurized activated carbon at outlet |
Temperature sensors |
Pt-100 with 0.01 °C accuracy |
Flowmeter |
electronic mass flow controller (MFC) |
Connectors |
for tubing 4mm i. d. / 6mm o. d. |
TECHNICAL DATA |
Power supply |
230 VAC / 50 Hz (110 VAC / 60 Hz option) |
Power consumption |
30 VA |
Dimensions (W x H x D) |
45 x 18 x 27 cm |
Weight |
approx. 5 kg |
APPROVALS & CERTIFICATIONS
- Our products are manufactured according to the ISO 9001 quality regulations.
- CE approved
Documentation
MC_3000_generation_mercury_calibration_gas_EN_1019